Do we really need plain and soft-tissue radiographies to detect radiolucent foreign bodies in the ED?
Original Contribution
Do we really need plain and soft-tissue radiographies to detect radiolucent foreign bodies in the ED?
Ibrahim Turkcuer MDa, Ridvan Atilla MDa,*, Hakan Topacoglu MDa, Sedat Yanturali MDa, Selahattin Kiyan MDa, Neslihan Kabakci MDb,
Seyran Bozkurt MDa, Arif Alper Cevik MDc
aDepartment of Emergency Medicine, Dokuz Eylul University Hospital, Inciralti, 35340 Izmir, Turkey bDepartment of Radiology, Dokuz Eylul University Hospital, Inciralti, 35340 Izmir, Turkey cDepartment of Emergency Medicine, Osmangazi University Hospital, 26480 Eskisehir, Turkey
Received 27 January 2006; revised 28 February 2006; accepted 16 March 2006
Abstract
Objective: The objective of this study was to compare 3 Imaging techniques—Plain radiography, soft- tissue radiography, and ultrasonography-in detecting nonradiopaque foreign bodies in soft tissue.
Methods: In this randomized, blinded, and descriptive in vitro study, 40 chicken thighs with 2 types of nonradiopaque foreign bodies (wood and rubber) and 40 chicken thighs as part of a control group were evaluated to detect soft-tissue foreign bodies with plain radiography, soft-tissue radiography, and high- frequency ultrasonography.
Results: The overall sensitivity, specificity, as well as positive predictive and negative predictive values of plain radiography for both nonradiopaque foreign bodies were 5%, 90%, 33%, and 48%, respectively; those of soft-tissue radiography for both nonradiopaque foreign bodies were 5%, 90%, 33%, and 48%, respectively; and those of ultrasonography for both nonradiopaque foreign bodies were 90%, 80%, 81%, and 89%, respectively.
Conclusions: In this Experimental model, the results show that high-frequency ultrasonography is superior to plain and soft-tissue radiographies and that the latter 2 techniques are similarly poor at detecting nonradiopaque foreign bodies.
D 2006
Objective
Various types of foreign bodies become embedded in soft tissue and can be difficult to detect in clinical practice [1,2]. If the interior of a wound is incompletely visualized, then
This article was presented as a poster in the Third Mediterranean Emergency Medicine Congress, September 1-5, 2005, Nice, France.
* Corresponding author. Tel.: +90 232 412 27 02; fax: +90 232 259 97 23.
E-mail address: [email protected] (R. Atilla).
further diagnostic tests are often justified in the ED to exclude foreign body presence. Plain radiography is the customary initial screening modality for a suspected foreign body [3]. It is well known that radiopaque bodies such as metal, glass, and gravel are easily detected on plain radio- graphs, but detection of nonradiopaque (radiolucent) frag- ments such as rubber, thorn, and wood is problematic [4].
Ultrasonography has an increasing role in diagnosing radiolucent foreign bodies in the ED. However, the diagnostic relevance of ultrasonography is related to its
0735-6757/$ - see front matter D 2006 doi:10.1016/j.ajem.2006.03.013
technical properties, as the length of the wavelengths (frequency) produced by transducers determines the depth of examination area into the tissue.
Primary applications of emergency ultrasound include trauma ultrasound, emergency ultrasound in pregnancy, emergency echocardiography, biliary and renal ultra- sound, and Foreign body localization during procedural ultrasound [5,6].
Because the primary clinical applications of ultrasonog- raphy are frequently on intra-abdominal organs or Cardiac abnormalities in the ED, low-frequency transducers (3.5- 5 MHz) are most likely to be available in the ED. Unfortunately, these are clearly suboptimal for foreign bodies [7,8].
There are several studies with poor sensitivity and specificity rates of ultrasonography, so other imaging modalities such as soft-tissue radiography, computed tomography, and magnetic resonance imaging (despite their limitations such as availability and cost-effective- ness) have been suggested to detect foreign bodies in soft tissue [9-13].
To date, however, the accuracy of higher-frequency ultrasonography has not been compared with plain and soft-tissue radiographies.
The purpose of this study was to investigate the comparative accuracy of high-frequency transducer ultraso- nography as well as plain and soft-tissue radiographies in determining the nonradiopaque soft-tissue foreign body in an established chicken thigh model, with the hypothesis that high-frequency ultrasonography could eliminate the option of plain and soft-tissue radiographies in the ED.
Methods
In this randomized, blinded, and descriptive study, 40 chicken thighs were evaluated for nonradiopaque foreign bodies with standard (plain) 2-view radiography, soft-tissue radiography, and 12.5-MHz ultrasonography.
Fig. 1 Plain radiography image of wood.
Fig. 2 Plain radiography image of rubber.
Study setting
-
-
- Foreign body acquisition and preparation: 20 pieces of each type of material-wood (toothpick) and rubber (shoe sole)-to represent 2 types of nonradiopaque foreign bodies were prepared. Wooden pieces were prepared in 5 mm of length and 1 mm in diameter from toothpicks, and rubber pieces were cropped in 5 x 5 x 2 mm in size from a used sport shoe’s sole.
- Soft-tissue model and chicken thigh acquisition: chicken thighs are preferred because of their proximity to the hand and foot anatomy in puncture-wound traumas. Eighty chicken thighs were taken daily from a chicken fresh cutting market and kept cool in a refrigerator at 48C. All the chicken thighs were of similar weight (mean = 250 F 45 g) and size.
- Chicken thigh model preparation: 40 foreign bodies were implanted in 40 fresh chicken thighs; 40 other chicken thighs were used as part of a control group. A similar technique was applied to all chicken thighs to simulate an open puncture wound: with the use of a no.15 scalpel, 10-mm skin incisions were applied with 308 of stabbing and a hemostat was inserted in a 1-cm depth into meat and manipulated inside the meat. One foreign body had been embedded in each chicken thigh and manipulated with hemostats to ensure uniform tissue damage. Other 40 chicken thighs were used as part of a control group, with similar incisions and manipulations with hemostats applied without inserting any foreign body.
- Diagnostic imaging: images were obtained using plain radiography, soft-tissue radiography, or ultrasonogra- phy. plain radiographs were of 2 views (anteroposterior [AP] and lateral) with standard dosage (48 kV and 2.5 mA for AP view; 52 kV and 2.7 mA for lateral view). Soft-tissue radiographs were of 2 views (AP and lateral) but with a low kilovolt dosage (45 kV and 2.8 mA for AP view; 50 kV and 3.2 mA for lateral view). Ultrasonography images were of 2 views (transverse and longitudinal) with a high-frequency linear trans-
-
Fig. 3 Soft-tissue radiography image of wood.
ducer (12.5 MHz; ATL 5000 HDI, Bathell, Wash) without a standoff pad.
-
-
- Diagnostic image interpretation: 2 attending radiolog- ists with at least 5 years of professional experience were enlisted. They were blinded to each other’s interpretations and model preparation.
-
Data analysis
The Statistical Package for Social Sciences (version 11.0, SPSS, Chicago, Ill) for Windows was used for statistical evaluation within 95% confidence intervals. The sensitivity, specificity, positive likelihood ratio, negative likelihood ratio, negative predictive value, and positive predictive value of each technique were calculated.
Results
Plain radiography
The radiologists failed to detect any of the wood foreign bodies in 20 chicken thighs (0%) (Fig. 1). In the
Fig. 4 Soft-tissue radiography image of rubber.
Fig. 5 Ultrasonography image of wood.
control group, 2 false-positive wood foreign bodies were detected by the radiologists (the sensitivity, specificity, as well as positive predictive and negative predictive values were 0%, 90%, 0%, and 47%, respectively). The positive likelihood ratio and negative likelihood ratio were 0 and 1.1, respectively.
With the plain radiography technique, only 2 of 20 rubber foreign bodies (10%) were detected (Fig. 2). In the control group, 2 false-positive rubber foreign bodies were also detected by the radiologists (the sensitivity, specificity, as well as positive predictive and negative predictive values were 10%, 90%, 50%, and 50%, respectively). The positive likelihood ratio and negative likelihood ratio were 1 and 1, respectively.
Soft-tissue radiography
The results of soft-tissue radiography were not different from those of plain radiography. The radiologists failed to detect any of the wood foreign bodies in 20 chicken thighs (0%) (Fig. 3). In the control group, 2 false-positive wood
Fig. 6 Ultrasonography image of rubber. Long arrows, image of toothpack; short arrow, shadow of the image.
Table 1 Overall sensitivity, specificity, positive predictive and negative predictive values, as well as positive and negative likelihood ratios of the 3 modalities used for both wood and rubber foreign bodies |
||||||
Sensitivity (%) |
Specificity (%) |
PPV (%) |
NPV (%) |
LR positive |
LR negative |
|
5 |
90 |
33 |
48 |
0.5 |
1.05 |
|
Soft-tissue radiography |
5 |
90 |
33 |
48 |
0.5 |
1.05 |
Ultrasonography |
90 |
80 |
81 |
89 |
4.5 |
0.5 |
PPV indicates positive predictive value; NPV, negative predictive value; LR, likelihood ratio. |
foreign bodies were detected by the radiologists (the sensitivity, specificity, as well as positive predictive and negative predictive values were 0%, 90%, 0%, and 47%, respectively). The positive likelihood ratio and negative likelihood ratio were 0 and 1.1, respectively.
Only 2 of 20 rubber foreign bodies (10%) in 20 chicken thighs were detected with the soft-tissue technique (Fig. 4). In the control group, 2 false-positive rubber foreign bodies were also detected by the radiologists (the sensitivity, specificity, as well as positive predictive and negative predictive values were 10%, 90%, 50%, and 50%, respec- tively). The positive likelihood ratio and negative likelihood ratio were 1 and 1, respectively.
Ultrasonography
Seventeen of 20 wood foreign bodies (85%) were detected positive with ultrasonography in 20 chicken thighs, whereas 4 false-negative wood foreign bodies were detected in the control group (the sensitivity, specificity, as well as positive predictive and negative predictive values were 85%, 80%, 81%, and 83%, respectively) (Fig. 5). The positive likelihood ratio and negative likelihood ratio were
4.25 and 0.19, respectively.
Nineteen of 20 rubber foreign bodies (95%) were detected by ultrasonography in 20 chicken thighs (Fig. 6). The number of false-positive rubber foreign bodies was also 4 in the control group (the sensitivity, specificity, as well as positive predictive and negative predictive values were 95%, 80%, 82%, and 94%, respectively). The positive likelihood ratio and negative likelihood ratio were 4.75 and 0.06, respectively.
For nonradiopaque foreign bodies, the overall sensitivity, specificity, positive likelihood ratio, negative likelihood ratio, positive likelihood ratio, negative likelihood ratio, as well as positive predictive and negative predictive values of the 3 modalities used are shown in Table 1.
Discussion
In this in vitro study, we compared plain radiography, soft-tissue radiography, and high-frequency ultrasonography in detecting foreign bodies in soft tissue. Our results revealed that using a higher-frequency ultrasonography to detect nonradiopaque foreign bodies could eliminate the inessential use of plain and soft-tissue radiographies.
The evaluation of a soft-tissue foreign body typically begins with routine radiography if the physical examination fails to uncover the suspected foreign body.
In the literature, it is well documented that plain radiographs are successful in detecting radiopaque foreign bodies in soft tissue [3]. However, the detection of nonradiopaque foreign bodies in the soft tissue limits the utility of relying on plain radiographs in all cases [14-17].
The poor accuracy of plain radiography has led some authors to conclude that soft-tissue radiographs could better reveal details through the soft tissue and could be successful in detecting foreign bodies [10,11,18]. Soft-tissue radiogra- phy is a radiographic modality in which the technique factors are adjusted to enhance and display the anatomical details of soft tissue. In the soft-tissue technique, decreasing the x-ray tube voltage lightens the film and enhances the contrast between some materials and the surrounding tissue. Altering Exposure factors as kilovolts, current, and time may discriminate between the densities of Soft tissues and foreign bodies [19].
However, none of these suggestions has been corrobo- rated with clinical studies in the medical literature. In a case report, Koornhof et al [20] concluded that especially the presence and the location of a swallowed fish bone could easily be detected with lateral soft-tissue radiography. However, it was a virtual case report that externally taped fishbone to one’s throat and used soft-tissue radiography. In one clinical observational study in which most patients with a complaint of foreign body sensation had undergone soft- tissue radiography of the neck, it was concluded that routine use of radiography in the assessment of foreign body in the throat was inappropriate [21].
In our animal study, plain and soft-tissue radiographies failed to detect nonradiopaque foreign bodies in soft tissue. Although a low kilovolt dosage was used for soft-tissue radiography according to plain radiography (45 vs 48 kV for AP views; 50 vs 52 kV for lateral views), the sensitivity and specificity of both modalities were the same. Soft-tissue radiography was as accurate as plain radiography in our study, and altering the x-ray dosage did not help physicians detect the nonradiopaque foreign body. (We applied an approximately 6% reduction in kilovolts for a low dosage according to the standard dosage to accommodate the smaller thickness of chicken thighs compared with actual human tissue, but this should be investigated with further studies).
Several studies have tested the effectiveness of ultraso- nography in detecting nonradiopaque foreign bodies in soft tissue. Investigations with high-frequency transducers had better results than those with low-frequency transducers [22-26], suggesting that transducer frequency and operator competence are limiting factors in clinical practice. The power of ultrasonography is important as the depth of penetration of the ultrasound wave into the body is directly related to wavelength (shorter wavelengths [higher frequen- cy] penetrate a shorter distance than longer wavelengths [lower frequency]). The higher the frequency used, the better the resolution (the ability to distinguish 2 adjacent objects). When the frequency is increased, more of the ultrasound beam is absorbed by the medium and the beam penetration declines. (The penetration depth into tissue and axial resolution are approximately 10 cm and 1 mm for
3.5 MHz, approximately 7 cm and 0.6 mm for 5 MHz, approximately 5 cm and 0.4 mm for 7.5 MHz, and approximately 2 to 20 cm and 0.2 mm for 12 MHz,
respectively [27].
Overall sensitivities and specificities with a 7.5-MHz transducer ultrasonography in detecting nonradiopaque foreign bodies are imperfect, but many still felt that ultrasound was possibly useful in screening for superficial foreign bodies [4,7]. Since the introduction of higher- frequency transducers, more accurate images have been obtained [17,22-24,27]. We conducted our research with a 12.5-MHz frequency transducer and found higher-frequency ultrasonography superior not only to plain and soft tissue radiographies but also to low-frequency ultrasonography. Although we have used one type of transducer in this study, when we compared our results with those of other clinical studies, we concluded that higher-frequency ultrasonogra- phy is superior to low-frequency transducers [17,22-24]. (It should be noticed that the efficiency of a 12.5-MHz ultrasonography is limited to superficial foreign bodies embedded less than 2 cm in depth of soft tissue, but there is no research specifically revealing the average depth of foreign body in soft tissue).
Other modalities have been suggested to detect soft- tissue foreign bodies. If a nonradiopaque object cannot be visualized on plain radiographs, alternative imaging techni- ques such as computed tomography and magnetic resonance imaging have been suggested; however, technical problems, cost-effectiveness, the time required to complete the test, and availability in clinical settings limit widespread use of these methods in the ED [28-31]. (Xeroradiography is no longer suggested as an alternative method [9].) New methods such as diffraction-enhanced imaging and multi- ple-image radiography show promise in visualizing a variety of soft tissues [32,33].
Wood and rubber were used as nonradiopaque foreign bodies in this study. Wood is the most common non- radiopaque foreign body in soft tissue, and plain radio- graphs have been reported to reveal a wooden foreign body in only 15% of patients [1]. Rubber is also another common
foreign body type in penetrating foot injuries. (When a nail penetrates shoes, socks, and feet, it may drive rubber into the wound cavity.) Sport shoes with rubber soles especially predispose to infection with Pseudomonas aeruginosa after penetrating wounds [2,34].
There are several limitations to this study. First of all, we could not correlate the interobserver reliability because only 2 radiologists individually interpreted the results of ultraso- nography and x-ray views. personal experiences and skills of the radiologists may play a role in the interpretation of the results; however, a previous study demonstrated a strong concordance between 2 radiologists in detecting foreign bodies with ultrasonography [35]. Second, this was an established chicken thigh model. Although the composition of layers of soft tissue in a chicken thigh is similar to that of human soft tissue, the thickness of the layers may be different and may cause altered absorption and reflection values of ultrasound waves and x-rays. Third, the depth of the ultrasound waves with a 12.5-MHz frequency transducer is limited to 2 to 20 mm, so deeply embedded foreign objects could have been missed.
We conclude that sonography is superior to plain and soft-tissue radiographies and that the latter 2 techniques are both poor at detecting nonradiopaque foreign bodies. If the type of foreign body is felt to be nonradiopaque and less than 2 cm deep, high-frequency ultrasonography might be the logical choice. In the clinical setting, an effort to detect the foreign body with plain and soft-tissue radiographies before ultrasonography will result in increased radiation exposure as well as increased cost and probably in a waste of time without significant clinical benefit. We also suggest that plain and soft-tissue radiography series should no longer be used as diagnostic imaging modalities to detect nonradiopaque foreign bodies if high-frequency ultrasonog- raphy is available in the ED.
References
- Anderson MA, Newmeyer III WL, Kilgore Jr ES. Diagnosis and treatment of retained foreign bodies in the hand. Am J Surg 1982;144: 63 - 7.
- Chang HC, Verhoeven W, Chay WM. Rubber foreign bodies in puncture wounds of the foot in patients wearing rubber-soled shoes. Foot Ankle Int 2001;22:409 - 14.
- Flom LL, Ellis GL. Radiologic evaluation of foreign bodies. Emerg Med Clin North Am 1992;10:163 - 77.
- Manthey DE, Storrow AB, Milbourn JM, et al. Ultrasound versus radiography in the detection of soft-tissue foreign bodies. Ann Emerg Med 1996;28:7 - 9.
- American College of Emergency Physicians. Council resolution on ultrasound. ACEP News; 1990.
- Mateer J, Plummer D, et al. Model curriculum for physician training in emergency ultrasonography. Ann Emerg Med 1994;23:95 - 102.
- Hill R, Conron R, Greissinger P, Heller M. Ultrasound for the detection of foreign bodies in human tissue. Ann Emerg Med 1997;29: 353 - 6.
- Banerjee B, Das RK. Sonographic detection of foreign bodies in the extremities. Br J Radiol 1991;64:107 - 12.
- Russell RC, Williamson DA, Sullivan JW, et al. Detection of foreign bodies in the hand. J Hand Surg 1991;16(A):2 - 11.
- Lammers RL, Magill T. Detection and management of foreign bodies in soft tissue. Emerg Med Clin North Am 1992;10:767 - 81.
- Ginsburg MJ, Ellis III GL. Detection of soft-tissue foreign bodies by plain radiography, xerography, computed tomography, and ultraso- nography. Ann Emerg Med 1990;19:701 - 3.
- Klein B, McGahan JP. Thorn synovitis: CT diagnosis. J Comput Assist Tomogr 1985;9:1135 - 6.
- Oikarinen KS, Nieminen TM, Makarainen H, Pyhtinen J. Visibility of foreign bodies in soft tissue in plain radiographs, computed tomography, magnetic resonance imaging, and ultrasound. An in vitro study. Int J Oral Maxillofac Surg 1993;22:119 - 24.
- Kjhns LR, Borlaza GS, Seigel RS, et al. An in vitro comparison of computed tomography, xerography, and radiography in the detection of soft-tissue foreign bodies. Radiology 1979;132:218 - 9.
- Bodne D, Quinn SF, Cochran CF. Imaging foreign glass and wooden bodies of the extremities with CT and MR. J Comput Assist Tomogr 1988;12:608 - 11.
- Graham Jr DD. Ultrasound in the emergency department: detection of wooden foreign bodies in the soft tissues. J Emerg Med 2002;22: 75 - 9.
- Venter NG, Jamel N, Marques RG, et al. Evaluation of radiological methods for detection of wood foreign body in animal model [Article in Portuguese]. Acta Cir Bras 2005;20(Suppl 1):19 - 26.
- Mucci B, Stenhouse G. Soft tissue radiography for wooden foreign bodies-a worthwhile exercise? Injury 1985;16:402 - 4.
- Clark KC. Soft tissue. In: Grune Strotton, editor. Positioning in radiography, 7th ed. New York-London: Grune & Stratton; 1956.
p. 550 - 6.
- Koornhof PG, du Plessis LJ, Claassen AJ. Swallowed fish bones-is radiography justified? A case report. S Afr J Surg 1990;28:158 - 60.
- Gautam V, Phillips J, Bowmer H, Reichl M. Foreign body in the throat. J Accid Emerg Med 1994;11:113 - 5.
- Jacobson JA, Powell A, Craig JG, et al. Wooden foreign bodies in the soft tissue detection at US. Radiology 1998;206:45 - 8.
- Turner J, Wilde CH, Hughes KC, et al. Ultrasound-guided retrieval of small foreign objects in subcutaneous tissue. Ann Emerg Med 1997; 29:731 - 4.
- Gooding GA, Hardiman T, Sumers M, et al. Sonography of the hand and foot in foreign body detection. J Ultrasound Med 1987;6: 441 - 7.
- Blyme PJ, Lind T, Schantz K, et al. Ultrasonographic detection of foreign bodies in soft tissue. A human Cadaver study. Arch Orthop Trauma Surg 1990;110:24 - 5.
- Gilbert FJ, Campbell RS, Bayliss AP. The role of ultrasound in the detection of non-radiopaque foreign bodies. Clin Radiol 1990;41: 109 - 12.
- Otto C. Textbook of clinical echocardiography. 3rd ed. Philadelphia7 WB Saunders Co; 2004.
- Grechenig W, Clement HG, Grechenig M, et al. Ultrasound imaging of foreign bodies-an experimental study [Article in German]. Biomed Tech (Berl) 1996;41:308 - 15.
- Newman JT, Hunt DW. Toothpick foreign body in the foot. A case report. J Am Podiatr Med Assoc 1989;79:458 - 61.
- Taupmann RE, Martin JE. The detection of nonopaque foreign bodies by xeroradiography. South Med J 1975;68:1186 - 7.
- Donaldson JS. radiographic imaging of foreign bodies in the hand. Hand Clin 1991;7:125 - 34.
- Li J, Zhong Z, Lidtke R, et al. Radiography of soft tissue of the foot and ankle with diffraction enhanced imaging. J Anat 2003;202:463 - 70.
- Muehleman C, Li J, Zhong Z, et al. Multiple-image radiography for human soft tissue. J Anat 2006;208:115 - 24.
- Patzakis MJ, Wilkins J, Brien WW, Carter VS. Wound site as a predictor of complications following deep nail punctures to the foot. West J Med 1989;150:545 - 7.
- Fornage BD, Schernberg FL. Sonographic diagnosis of foreign bodies in distal extremities. Am J Radiol 1986;147:567 - 9.