Article

Evaluation of gum-elastic bougie combined with direct and indirect laryngoscopes in vomitus setting: A randomized simulation trial

a b s t r a c t

Purpose: Videolaryngoscopes may not be useful in the presence of vomitus due to blurred images on the monitor. The objective of our study is to compare the utility of gum-elastic bougie application for tracheal intubation with the Macintosh laryngoscope (McL), which is a direct laryngoscope, with that of the Pentax-AWS Airwayscope(R) (AWS) and McGRATH(R) MAC (McGRATH) in simulated vomitUS settings.

Methods: Sixteen novice doctors performed tracheal intubation on an adult manikin using McL, AWS, and McGRATH with or without GEB under normal and vomitus simulations.

Results: In the normal setting the tracheal intubation was successful with the three laryngoscopes regardless of GEB application. In the vomitus setting, the intubation success rate did not significantly improve using McL, while it did using McGRATH or AWS. In the normal settings, GEB application significantly lengthened the intuba- tion time in all three laryngoscopes. By contrast, in the vomitus settings, GEB application significantly shortened the intubation time in all three laryngoscopes. For the comparison of three laryngoscopes, the intubation time did not differ significantly in normal setting, while it was significantly longer in McG and AWS trials than McL trial. Conclusion: The GEB application facilitates the tracheal intubation in the vomitus setting using McGRATH and AWS in adult simulation.

(C) 2016

  1. Introduction

The Pentax Airwayscope(R) (AWS; Hoya, Tokyo, Japan) is a videolaryngoscope reported to provide an indirect view of the airway [1,2]. Studies indicate that AWS is useful not only for difficult airway management but also for emergent tracheal intubation during resusci- tation by simulation analysis [3,4]. The McGRATH(R) MAC (McGRATH; Aircraft Medical Ltd., Edinburgh, UK) is a device that has been developed with a high-resolution video camera, providing direct and indirect views of the glottis and reportedly useful for intubating several difficult airways [5,6]. While AWS and McGRATH are both considered conve- nient tools for difficult or emergent airway management, we have pre- viously reported that their utility is inhibited in the presence of vomitus in the pharynx [7].

The gum-elastic bougie (GEB), a tracheal tube introducer, is com- monly used in airway management and its use is recommended by var- ious guidelines at early stages of difficult intubation [8]. Several studies have been published regarding its utility in the context of difficult adult airway management, particularly for addressing difficult laryngoscopy

* Corresponding author at: Department of Anesthesiology, Osaka Medical College, Daigaku-machi 2-7, Takatsuki, Osaka 569-8686, Japan.

E-mail address: [email protected] (N. Komasawa).

situations. Furthermore, evaluations of GEB application during chest compressions have been reported [9,10].

Application of GEB for tracheal intubation with direct (McL) and in- direct laryngoscopes (AWS and McGRATH) for tracheal intubation in the vomitus setting has not yet been validated; therefore, we decided to compare the utility of GEB application for tracheal intubation with McL, AWS, and McGRATH in vomitus settings. Because direct clinical evaluation is unethical, we hypothesized that GEB application may facil- itate the tracheal intubation with the three laryngoscopes in vomitus settings using an adult manikin with vomitus simulations.

  1. Methods

From April to July 2016, 16 novice doctors performed trials after completing one month of anesthesia training. Written informed consent was obtained before the study. This study was approved by the Osaka Medical College Research Ethics Committee (Approval number 1321).

The Airway Trainer(R) (Laerdal, Stavanger, Norway), designed to ac- curately represent an adult male, was used for the study simulations and intubations. We used a 15Fr gum-elastic bougie (Tracheal tube in- troducer(TM), Portex, St. Paul, MN, USA) in this study. Tracheal tubes (Portex, St. Paul, MN, USA) with an internal diameter of 7.5 mm were used [11].

http://dx.doi.org/10.1016/j.ajem.2016.12.032

0735-6757/(C) 2016

VAS scores for difficulty of trache”>Simulated stomach contents (vomitus; Simulated stomach con- tents(R), Laerdal, Norway) were added to the pharynx of the manikin. The contents were prepared by dissolving 10 g of powder in 200 mL of water according to the manufacturer’s instructions and were poured into the pharynx to the level of covering the epiglottis to simulate vom- itus or hematemesis [7]. The lower esophagus was clamped with for- ceps to keep these liquids in the pharynx. We also clamped the both bronchus instead of the trachea because clamping the trachea impedes smooth tracheal intubation.

The manikin was placed on a hard, flat table for “on the resuscitation bed” simulation. Each participant was instructed to insert the tracheal tube with the three laryngoscopes (McL, AWS, McGRATH), attach a bag valve mask, and attempt to ventilate the lungs of the manikin. In McL and McGRATH trials, participants used size 4 blade. In AWS trial, standard Intlock(R) (ITL-SL, HOYA, Japan) was used. Participants were given 10 min to practice intubation, with the instructor available for ad- vice. The appropriate equipment for each trial was placed in a box next to the manikin’s head. Intubation started when the participant picked up McL, AWS, or McGRATH and ended at the point of manual ventilation after tube insertion [12]. The number of intubation sessions was record- ed for both tracheal and Esophageal intubations. At the end of the study, participants rated the difficulty of using each device using a visual ana- log scale (VAS) from 0 mm (extremely easy) to 100 mm (extremely dif- ficult) [13].

Statistical analysis was performed utilizing JMP(R) 11 (SAS Institute

Inc., Cary, NC, USA). Results obtained from each trial were compared using one-way repeated measures analysis of variance for intubation time and VAS and Fisher’s exact test for the success rate. Data are pre- sented as means +- standard deviations (SDs). A P-value of b 0.05 was considered statistically significant.

The study was designed as a randomized crossover trial to minimize

the order effect. In each McL, AWS, and McGRATH trial, participants per- formed tracheal intubation in both simulations (normal and vomitus). The order of intervention was randomized for each participant using the random number table, resulting in a total of six interventions per participant.

The sample size was calculated on the basis of our preliminary study on the time required for intubation with McL and McGRATH in the vom- itus setting in eight participants. The mean (SD) time was 13.3 +- 3.2 s for McL and 6.9 +- 2.5 s for McGRATH. We considered that 5 s is a signif- icant difference. Using an alpha error of 0.05 and a beta error of 0.2, we estimated that 14 participants would be adequate for each group. Therefore, we planned to recruit 16 participants for each group to adjust for missing data.

  1. Results

The number of times the participants had worked before participat- ing in the trial with McL, AWS, and McGRATH were 24.1 +- 5.8 s, 3.7 +-

1.6 s, and 2.4 +- 1.2 s, respectively.

Intubation success with or without GEB using McL, AWS, and McGRATH

The number of successful tracheal intubations for each device is displayed in Table 1. In the normal setting the tracheal intubation was successful with the three laryngoscopes regardless of GEB application (Table 1a). In the vomitus setting, the intubation success rate did not significantly improve using McL, while it did using McGRATH or AWS (Table 1b).

Intubation time with or without GEB using McL, AWS, and McGRATH

The intubation time in each setting is shown in Fig. 1. In the normal setting, GEB application significantly lengthened the intubation time in all three laryngoscope trials. In contrast, the GEB application significant- ly shortened the intubation time in all three laryngoscopes.

Table 1

Tracheal intubation success rates for McL, AWS, and McGRATH in normal (a) and vomitus

(b) settings. Numerator: number of participants who were successfully intubated. Denom- inator: number of participants for whom tracheal intubation was attempted. *P b 0.05 compared to without GEB, #P b 0.05 compared to McGRATH and McL.

Without GEB (successful/total)

With GEB (successful/total)

P-value

(a)

McL

16/16

16/16

1.00

AWS

16/16

16/16

1.00

McGRATH

16/16

16/16

1.00

(b) McL

12/16

16/16

0.101

AWS

8/16

15/16

0.016*

McGRATH

11/16

16/16

0.043*

AWS, Pentax-AWS Airwayscope(R); McGRATH, McGRATH(R) MAC; McL, Macintosh laryngoscope.

For the comparison among the three laryngoscopes, the intubation time did not differ significantly in normal setting, while it was signifi- cantly longer in McG and AWS trials than McL trial.

VAS scores for difficulty of tracheal intubation with McL, AWS, and McGRATH

As shown in Fig. 2, the subjective difficulty of laryngoscopy tracheal intubation did not differ regardless of GEB application in the normal set- ting, while it was significantly lowered by GEB application using all three laryngoscopes in the vomitus setting.

In the vomitus setting, the VAS without GEB was significantly higher in the AWS trial than in the McL and McGRATH trials.

  1. Discussion

Airway management is considered an essential element, particularly for In-hospital CPR. While conventional direct-view laryngoscopes such as McL are the most widely used for tracheal intubation, it is difficult to master the skills required for use, and the incidence of inaccurate intu- bation can be unacceptably high for occasional operators [14,15]. A major problem encountered during airway management is vomitus or blood in the pharynx and neck fixation with the cervical collar. A large number of patients exhibit vomiting or hematemesis during sudden cardiac arrest, leading to difficulty in tracheal intubation during resusci- tation [16].

Increasing evidence indicates that videolaryngscope is suitable for tracheal intubation during various difficult airway management and emergency situations [17,18]. However, one clinical study showed that AWS did not show superiority to McL in prehospital settings, as op- posed to simulated in-hospital situations [19]. We speculate that vomi- tus or hematemesis may have contributed to the lower success rate of AWS in the prehospital situations. We previously showed the inferiority of AWS to McL in the vomitus setting.

The GEB is a commonly used airway adjunct in intubation and rec- ommended by several guidelines for use at an early stage in cases of dif- ficult tracheal intubation. Evidence from adult patients suggests that doctors can secure the airway with a high success rate when using the GEB [10,20]. In the previous study, we found that intubation time was significantly longer with chest compressions, regardless of GEB use, al- though the intubation success rate was significantly higher with the GEB. The subjective difficulty for tube passage through the glottis was also lower with the GEB than without during chest compressions.

The present study found that the intubation time with AWS was sig- nificantly shortened by GEB application than in the vomitus setting, ac- companied by a significant intubation success rate difference. Furthermore, the intubation time was significantly shortened by

Fig. 1. Box-and-whisker plot (median, IQR, and range) of time required for simulated tracheal intubation in normal and vomitus settings using each laryngoscope. (a) Normal setting, (b) vomitus setting. AWS, Pentax-AWS Airwayscope(R); McGRATH, McGRATH(R) MAC; McL, Macintosh laryngoscope. *P b 0.05 compared to without GEB, #P b 0.05 compared to McGRATH and McL.

a)

a)

application in McG and McL trials. One possible reason for this is that novice doctors can sense the “clicks” and “hold up” of the GEB even in vomitus setting and perform definite tracheal intubation with GEB assistance.

Recently, preload bougie technique has been reported [21]. A poten- tially more advantageous preloaded bougie-assisted intubation tech- nique has been demonstrated in a YouTube video demonstration in a cadaver [22]. With this technique, when “hold up” is felt, the tip of the tracheal tube is already positioned right in front of the glottis and with one continuous fluid motion the tube together with the bougie is then rotated 90? clockwise so that the side bevel tip is oriented such that it does not get hung up on the right arytenoid cartilage while advancing the tracheal tube off of the bougie and into the trachea. With the other technique, intubation is not performed in one continuous fluid motion, and when “hold up” is felt, the tracheal tube has to be advanced before the tip reaches the glottis. Evaluation of this preload bougie technique in the vomitus settings is warranted in the future.

  1. Limitations

This study has several limitations. First, intubation was performed on a manikin, which leads to shorter airway intervention times than those required for actual patients. Second, the use of these three devices may be less than ideal in patients with Difficult airways, such as those with a severely Restricted mouth opening or a small jaw. Third, the sim- ulations do not account for other factors related to resuscitation, such as

chest compression and cervical stabilization [23,24]. Finally, the homo- geneity of vomitus cannot completely simulate clinical situations.

  1. Conclusion

Within the limitations of our study, we conclude McGRATH and McL can show superior performance compared to AWS for tracheal intuba- tion in adults with vomitus or blood in the pharynx. Future studies ex- amining accumulated evidence of clinical experiences and randomized trials of McL, AWS, and McGRATH in actual patients with hematemesis or vomitus are required to clarify the findings of this simulation study.

Competing interests

The authors have no affiliation with the manufacturers of any of the devices described in the manuscript and declare no financial interest in the material described in the manuscript. Financial support for the study was provided by our institution and department.

Author contributions

F.O., N.K., R.M., and K.H. were involved in the study design, study im- plementation, data analysis, and manuscript preparation. T.M. was in- volved in the study design and manuscript preparation.

Fig. 2. Box-and-whisker plot (median, IQR, and range) of visual analog scale scores for difficulty of simulated tracheal intubation in normal and vomitus settings using each laryngoscope.

a)

b)

(a) Normal setting, (b) vomitus setting. AWS, Pentax-AWS Airwayscope(R); McGRATH, McGRATH(R) MAC; McL, Macintosh laryngoscope. *P b 0.05 compared to without GEB, #P b 0.05 compared to McGRATH and McL.

References

  1. Neumar RW, Shuster M, Callaway CW, Gent LM, Atkins DL, Bhanji F, et al. Executive summary: 2015 American Heart Association Guidelines Update for cardiopulmonary resuscitation and emergency cardiovascular care. Circulation 2015;132:S315-67.
  2. Ohchi F, Komasawa N, Imagawa K, Okamoto K, Minami T. Evaluation of the efficacy of six supraglottic devices for airway management in dark conditions: a crossover randomized simulation trial. J Anesth 2015;29:887-92.
  3. Asai T, Liu EH, Matsumoto S, Hirabayashi Y, Seo N, Suzuki A, et al. Use of the Pentax- AWS in 293 patients with difficult airways. Anesthesiology 2009;110:898-904.
  4. Komasawa N, Kido H, Miyazaki Y, Tatsumi S, Minami T. Cricoid pressure impedes

tracheal intubation with the Pentax-AWS Airwayscope(R): a prospective randomized trial. Br J Anaesth 2016;116:413-6.

  1. Komasawa N, Fujiwara S, Mihara R, Minami T. Comparison of McGrath(R) and Pentax- AWS Airwayscope(R) for tracheal intubation by anaesthesiologists during chest com- pression in a manikin: a randomized crossover trial. Eur J Anaesth 2015;63:582-5.
  2. Deguchi S, Komasawa N, Kido H, Ueno T, Minami T. Impact of pillow height on dou- ble-lumen endotracheal tube intubation with McGRATH(R) MAC: a prospective ran- domized clinical trial. J Clin Anesth 2016;34:339-43.
  3. Mihara R, Komasawa N, Matsunami S, Minami T. Comparison of direct and indirect laryngoscopes in vomitus and hematemesis settings: a randomized simulation trial. Biomed Res Int 2015;2015:806243.
  4. Henderson JJ, Popat MT, Latto IP, Peace AC, Difficult Airway Society. Difficult Airway Society guidelines for management of the unanticipated difficult intubation. Anaes- thesia 2004;59:675-94.
  5. Komasawa N, Hyoda A, Matsunami S, Majima N, Minami T. Utility of a gum-elastic bougie for difficult airway management in infants -a simulation-based crossover analysis. Biomed Res Int 2015;2015:617805.
  6. Komasawa N, Cho T, Mihara R, Minami T. Efficacy of gum-elastic budgie for tracheal intubation during chest compression in a manikin: a randomized crossover trial. Am J Emerg Med 2016;34:54-6.
  7. Okada D, Komasawa N, Fujiwara S, Minami T. Comparison of tube-guided and guideless videolaryngoscope for tracheal intubation during chest compression in a manikin: a randomized crossover trial. J Anesth 2015;29:331-7.
  8. Komasawa N, Ueki R, Itani M, Nishi S, Kaminoh Y. Validation of the Pentax-AWS air- way scope utility as an Intubation device during cardiopulmonary resuscitation on the ground. J Anesth 2010;24:582-6.
  9. Komasawa N, Ueki R, Yamamoto N, Nishi S, Kaminoh Y, Tashiro C. Comparison of Pentax-AWS Airwayscope, Airtraq and Miller laryngoscope for tracheal intubation by novice doctors during infant cardiopulmonary resuscitation simulation: a ran- domized crossover trial. J Anesth 2013;27:778-80.
  10. Katz SH, Falk JL. Misplaced Endotracheal tubes by paramedics in an urban emergen- cy medical services system. Ann Emerg Med 2001;37:32-7.
  11. Wang HE, Mann NC, Mears G, Jacobson K, Yealy DM. Out-of-hospital airway man- agement in the United States. Resuscitation 2011;82:378-85.
  12. Simons RW, Rea TD, Becker LJ, Eisenberg MS. The incidence and significance of emesis associated with out-of-hospital cardiac arrest. Resuscitation 2007;74: 427-31.
  13. Kohama H, Komasawa N, Ueki R, Samma A, Nakagawa M, Nishi S, et al. Comparison of Supreme(R) and Soft Seal(R) laryngeal masks for airway management during cardio- pulmonary resuscitation in novice doctors: a manikin study. J Anesth 2011;25: 98-103.
  14. Fujiwara S, Komasawa N, Matsunami S, Okada D, Minami T. Comparison of Pentax- AWS Airwayscope and Glidescope for infant tracheal intubation by anesthesiologists during cardiopulmonary arrest simulation: a randomized crossover trial. Biomed Res Int 2015;2015:190163.
  15. Arima T, Nagata O, Miura T, Ikeda K, Mizushima T, Takahashi A, et al. Comparative analysis of airway scope and Macintosh laryngoscope for intubation primarily for cardiac arrest in prehospital setting. Am J Emerg Med 2014;32:40-3.
  16. Cho T, Komasawa N, Hattori K, Mihara R, Minami T. Gum-elastic bougie efficacy for tracheal intubation during Continuous chest compression in infants-a crossover sim- ulation trial. J Emerg Med 2016;51:19-24.
  17. Baker JB, Maskell KF, Matlock AG, Walsh RM, Skinner CG. Comparison of preloaded bougie versus standard bougie technique for endotracheal intubation in a cadaveric model. West J Emerg Med 2015;16:588-93.
  18. Kevin High from Vanderbilt University. Take 5: 5 airway bougie tips. https://www. youtube.com/watch?v=TrHZAPLPKwU. [accessed on 16.12.10].
  19. Komasawa N, Ueki R, Kaminoh Y, Nishi S. Evaluation of chest compression effect on airway management with air-Q(R), aura-i(R), i-gel(R), and Fastrack(R) intubating

supraglottic devices by novice physicians: a randomized cross over simulation study. J Anesth 2014;28:676-80.

  1. Komasawa N, Ueki R, Kohama H, Nishi S, Kaminoh Y. Comparison of Pentax-AWS Airwayscope video laryngoscope, Airtraq optic laryngoscope, and Macintosh laryn- goscope during cardiopulmonary resuscitation under cervical stabilization: a mani- kin study. J Anesth 2011;25:898-903.

Leave a Reply

Your email address will not be published. Required fields are marked *